OIML R60 defines Y as (Emax − Emin) / vmin, where Emax and Emin are the upper and lower limits of the applicable measuring range and vmin is the minimum load-cell verification interval.
The two terms connect a load cell’s resolution to its capacity, but they do not determine the performance of the complete weighing instrument.
What vmin and Y mean in OIML R60
OIML R 60-1:2021 defines Y as (Emax − Emin) / vmin, where Emax and Emin are the upper and lower limits of the applicable measuring range and vmin is the minimum load-cell verification interval. The earlier 2017 edition uses the same structure and notes that this ratio describes the resolution of the load cell independently of the load-cell test results [S4].
In practical notation, the relationship is Y = (Emax - Emin) / vmin [S6][S8]. Emax is the maximum capacity and Emin is the minimum load or lower measurement-range limit used in the applicable range definition. For a single-interval cell with Emin equal to zero, the expression reduces to Y = Emax / vmin.
This distinction matters: vmin is not a measured accuracy result. It is a declared metrological parameter used to describe the smallest verification interval supported by the cell.
Why the two ratios are related but not interchangeable
Both quantities use load and capacity, but they answer different engineering questions. vmin identifies the smallest verification interval in the applicable unit, while Y expresses how many such intervals span the defined maximum range [S5].
A numerical example makes the relationship explicit. If a cell has Emax = 1,000 kg, Emin = 0 kg and vmin = 1 kg, then Y = (1,000 - 0) / 1 = 1,000. If the same cell declares vmin = 2 kg, the ratio becomes (1,000 - 0) / 2 = 500. The maximum range has not changed, but the declared minimum verification interval has doubled.
The distinction prevents a common specification error. A large Y value does not by itself prove that the installed scale can resolve every corresponding interval. OIML R60 evaluates load-cell characteristics including combined error, creep, temperature effects, minimum dead-load output return and moisture behavior, while the indicator, mechanical structure and system configuration remain separate design considerations [S5].
Where the ratio applies in multi-interval instruments

OIML R60 defines Y as (Emax − Emin) / vmin, where Emax and Emin are the upper and lower limits of the applicable measuring range and vmin is the minimum load-cell verification interval. Each interval can have its own usable range and verification structure, so the applicable limits must be identified before calculating Y.
The load-cell definition provides the compact relation Y = (Emax - Emin) / vmin [S6]. Annex material applies the same equation and directs evaluators to inspect vmin, Y and Z for all cells of the same capacity [S8]. That inspection is important because identical nominal capacities do not establish that every installed cell has identical declared metrological parameters.
OIML R 60-1:2021 also contains an illustration section for its capacity, range and output definitions, covering 3.5, followed by measurement and error terms in 3.7 and abbreviations in 3.9 [S2]. The terminology must therefore be read with the associated range definitions, rather than treating “maximum capacity” as a substitute for every concept of measuring range.
How to compare cells with different vmin and Y values
For cell selection, vmin should be compared with the required division after accounting for the load assigned to each cell, while Y should be compared with the maximum range used in the calculation [S5]. Capacity alone does not establish suitability, and a higher accuracy-class number does not override platform, indicator or environmental limitations [S5].
Comparison should therefore follow four criteria: declared vmin in the required mass unit, Y for the applicable range, maximum capacity per cell, and the certificate limits for the intended operating conditions.
For example, a 2,000 kg cell with vmin = 2 kg gives Y = 1,000, while a 1,000 kg cell with vmin = 2 kg gives Y = 500. These cells have different capacities and different interval counts. If either is installed in a multi-cell platform, the load distribution and maximum load carried by each cell must be established before the class, vmin or Y is accepted into the system design [S5].
What vmin and Y do not prove

A manufacturer-specified vmin is not an accuracy class, and a load-cell certificate does not automatically mean that an installed scale has been approved. It is a load-cell parameter, while the complete weighing instrument also depends on the indicator, platform mechanics, excitation, wiring and load distribution.
The OIML R 60-1:2021 text contains metrological requirements in section 5 and technical requirements in section 6 [S2]. Its structure separates permissible measurement error, repeatability, reference-condition variation and influence quantities from the technical requirements for analogue-active and digital load cells. A vmin declaration should be considered together with the applicable accuracy class and other relevant metrological information.
Older OIML material also shows why the formulas require care. The 2000 edition erratum corrected the inequality governing vmin relative to the dead-load range and measurement number, including the expression vmin ≤ (Dmax - Dmin) / nmax [S3]. This is a further reason not to calculate Y from a catalogue headline or infer that one ratio alone establishes conformity.
Procurement checks for vmin and Y declarations
Request the manufacturer’s declared vmin, the associated minimum and maximum range values, and the resulting Y ratio for each proposed load cell. Confirm the units and verify whether the calculation uses Emax alone or the full expression (Emax - Emin) / vmin [S2][S6].
Then check the number of divisions against the actual platform design and the load assigned to each cell. A system using multiple cells requires their individual capacity, load distribution and minimum dead-load output return to be assessed together [S5]. Review the OIML certificate for class, capacity, temperature range and other stated limits rather than accepting a class code without its conditions [S5].
Track two signals when a cell or scale is re-specified: any change in vmin at unchanged Emax, and any change in Y caused by different Emin or Emax. Each change alters the resolution basis even when the accuracy class remains unchanged, while manufacturer and certificate documentation remain the controlling records [S1][S8].