How to Select an Accelerometer by mV/g Sensitivity

Accelerometer sensitivity directly sets the measurable range: a 10 mV/g IEPE sensor with ±5 V linear output reaches 500 gpk, while 100 mV/g reaches 50 gpk [S2].

The choice depends on peak acceleration, noise floor, frequency band and measurement purpose, with 10 and 100 mV/g serving as common vibration-testing sensitivities [S2][S3].

What mV/g sensitivity actually specifies

Sensitivity is the voltage change produced per unit of acceleration, and 1 g equals 9.81 m/s² [S8]. For an IEPE accelerometer specified at 100 mV/g, a 10 g signal produces 1,000 mV, or 1 V, at the sensor output [S5].

Nominal sensitivity is not the same as calibrated sensitivity. Calibration data should therefore control amplitude calculations instead of the catalogue value alone.

For acceleration expressed in m/s², a sensitivity of 9.81 mV/(m/s²) is numerically equivalent to 100 mV/g, and 0.981 mV/(m/s²) is equivalent to 10 mV/g. Charge-output accelerometers use a different unit, pC/g, rather than mV/g [S4][S8].

10 mV/g versus 100 mV/g

A 10 mV/g device provides ten times the voltage output of a 100 mV/g device for the same acceleration [S7]. That lower sensitivity extends headroom: with ±5 V output, the approximate limits are 500 gpk for 10 mV/g and 50 gpk for 100 mV/g [S2].

The trade-off is signal-to-noise ratio. Higher sensitivity produces a larger signal relative to electronic noise, while lower sensitivity is more vulnerable to the controller noise floor [S3]. A 100 mV/g sensor generally provides a higher signal-to-noise ratio, while a 10 mV/g sensor provides a higher measurement range for a given output limit [S2][S3].

Match sensitivity to the test level

vibration sensor sensitivity rated in mv per g - Match sensitivity to the test level
vibration sensor sensitivity rated in mv per g - Match sensitivity to the test level

Start with the highest credible peak acceleration, then calculate the required voltage as acceleration multiplied by sensitivity. A 20 g event measured by a 10 mV/g sensor requires 200 mV, while the same event measured by a 100 mV/g sensor requires 2,000 mV, or 2 V [S5][S7].

Most integrated-electronics piezoelectric (IEPE) accelerometers can output a ±5 V excitation signal [S2]. Leave signal-processing margin rather than calculating to an exact clipping point, especially for impact, transient or resonance measurements.

Low-level machines require the opposite calculation. If the controller cannot resolve the output of a 10 mV/g sensor cleanly, 100 mV/g can preserve low-amplitude detail. The selection procedure in Select Accelerometer Sensitivity for Slow, Low-Level Machines covers this low-level measurement problem in greater detail.

Frequency range changes the usable result

An individual accelerometer calibration chart can include a frequency-response curve [S4]. A 100 mV/g piezoelectric accelerometer with a 0.5–10,000 Hz range was used as a diagnostic reference in bearing testing [S1].

Frequency also affects charge-output behavior. Charge sensitivity is stated at 80 Hz and room temperature [S4]. IEPE cable length can also alter frequency response, which makes transducer installation part of the measurement chain rather than a secondary detail [S4].

For a broad selection framework covering frequency and mounting effects, see Select a Vibration Sensor for 1 Hz to 10 kHz. The accelerometer reference defines the sensor family, while a vibration sensor may use a different sensing technology and output format.

Noise, tolerance and cross-axis errors

vibration sensor sensitivity rated in mv per g - Noise, tolerance and cross-axis errors
vibration sensor sensitivity rated in mv per g - Noise, tolerance and cross-axis errors

Electronic noise may be specified across a broad bandwidth, such as 2–25,000 Hz, or at a discrete frequency in acceleration-equivalent units [S3]. Low-frequency machinery signals are particularly vulnerable because useful low-acceleration content often lies close to the noise floor.

Transverse sensitivity is another measurable error term. It compares output at 90° with output along the sensitive axis, and the transverse percentage equals maximum or average off-axis output divided by transducer sensitivity, multiplied by 100 [S3]. A sensor with less than 5% transverse response can still distort a result if the installation exposes it to significant lateral motion.

Metra states charge sensitivity at 80 Hz and room temperature [S4]. These conditions do not automatically apply across temperature, frequency and mounting changes, so field accuracy remains a system property.

Calibration and data-entry controls

Entering the wrong sensitivity can produce a tenfold amplitude error. The same error can distort shaker control and invalidate collected data.

TEDS reduces manual entry mistakes, but it does not replace the need to confirm that the selected sensor suits the test level.

For permanent machine monitoring, the condition monitoring system must also preserve the sensor’s scale. A vibration analyzer cannot recover a clipped peak or a signal buried below the acquisition noise floor merely through more accurate software processing.

Application fit and procurement record

vibration sensor sensitivity rated in mv per g - Application fit and procurement record
vibration sensor sensitivity rated in mv per g - Application fit and procurement record

Choose 10 mV/g when the expected range approaches hundreds of g and the controller has a sufficiently low noise floor. Choose 100 mV/g when resolving modest vibration levels is more important than measuring near 50 g. For shock applications, verify the manufacturer’s specified peak range rather than extrapolating indefinitely from nominal sensitivity [S2][S7].

A procurement record should state nominal sensitivity in mV/g, calibrated sensitivity, tolerance, frequency response, transverse sensitivity, excitation voltage, constant-current requirement, noise specification and mounting method. Individual calibration charts are valuable because they include measured sensitivity, transverse response, insulation resistance, IEPE bias voltage and frequency-response data [S4].

The traceable signals are sensor saturation at the intended peak level and a measurable drop in signal-to-noise ratio at the intended low level. Review both against the certified sensitivity before approving the installation.

Frequently asked questions

Which mV/g sensitivity should I choose for a vibration test reaching 500 gpk?

Choose a 10 mV/g IEPE accelerometer for a 500 gpk measurement with a ±5 V linear output range. At 10 mV/g, the available voltage headroom supports approximately 500 gpk, whereas a 100 mV/g sensor reaches about 50 gpk.

When is 100 mV/g preferable to 10 mV/g for low-level vibration?

Choose 100 mV/g when resolving modest vibration levels is more important than measuring near 50 g. Its higher voltage output improves signal relative to electronic noise, although its approximate ±5 V range limits peak acceleration to about 50 gpk.

How should I calculate the output voltage for a 20 g event?

Multiply the acceleration by the nominal sensitivity: a 20 g event produces 200 mV with a 10 mV/g sensor and 2,000 mV, or 2 V, with a 100 mV/g sensor. Use calibrated sensitivity for amplitude calculations rather than relying solely on the catalogue value.

What frequency-response and calibration checks are needed before specifying an accelerometer?

Confirm that the intended measurement band lies within the sensor’s specified frequency response; the usable upper limit is commonly the 3 dB point, generally around one-half of resonance. Also verify calibration conditions, frequency-dependent charge-sensitivity changes, and installation effects, including IEPE cable-length effects.

8 sources
  1. Evaluation of the Diagnostic Sensitivity of Digital Vibration ...
  2. Sensor Specifications - Vibration Control - VRU - VR University (Jun 10, 2022)
  3. Technical note
  4. Accelerometer Characteristics
  5. Measuring Vibration with Accelerometers
  6. Accelerometer Sensitivity and Output Units
  7. Transducer Sensitivity Selection for Vibration Control
  8. Accelerometer Specifications: Deciphering an ...

Need to source matching manufacturers or get a quote?

SpecForge connects industrial buyers with verified manufacturers. Submit your requirement and we will route it to matched suppliers.

Submit RFQ now →
Ask SpecForge AI