Resistance decade boxes use 11-position dials so a 10-step decade can overlap the next decade and change resistance with fewer abrupt transitions [S2][S3][S6].
Typical boxes use 3–11 rotary dials, with accuracy ranging from 0.01% to 1%, while the 11th position extends each dial beyond its normal digit range [S2][S3].
The 11th Position Is an Overlap, Not an Extra Decimal Digit
A conventional decade contains 10 resistor steps, numbered 0 through 9, with each step equal to the place value of that dial [S9]. An 11-position overlap dial adds position 10, which allows the selected section to carry the full tenfold value of that decade rather than resetting to zero.
For example, a units dial normally selects 0–9 Ω in 1 Ω increments, while the RZ-4500 uses eleven positions marked 0 through 10, so its units dial can select 0–10 Ω in 1 Ω steps [S4]. Its tens dial covers 0–100 Ω in 10 Ω steps, and each higher dial applies the next power-of-ten multiplier [S4].
This arrangement is not the same as reading a conventional decimal number. The overlap position deliberately represents one complete decade, which is why a dial can be labeled 0–10 even though ordinary decimal digit positions run from 0–9 [S3][S5][S7].
Overlap Reduces Abrupt Resistance Changes
Moving the hundreds dial from 9 to 10 changes 963 Ω directly to 1,063 Ω in one transition, whereas reducing the hundreds dial to zero and moving the higher decade from 0 to 1 changes 963 Ω to 63 Ω and then to 1,063 Ω, creating two sudden transitions [S4].
That distinction matters when the decade box is connected to a calibration target. The RZ-4500 documentation warns that sudden, large resistance changes can cause excessive and damaging surge currents in the device or system under calibration [S4]. The overlap position therefore gives the operator a controlled route through the boundary between two decades.
Guildline describes the same arrangement as an overlap 10 position that permits fine adjustment without resetting all dials when passing through a decade point [S6]. IET models also specify 11 contact studs and 10 resistors per decade, allowing adjacent dial values to overlap [S7].
Continuous Rotation Changes the Operating Method

The IET 1433 series has dials with an overlap 10 position [S5][S8].
The IET 1432 uses positive detent mechanisms and bar-type knobs so the operator can sense the switch position without looking at the panel [S7].
By contrast, some designs stop at position 10 to prevent an operator from accidentally switching directly from 10 to 0 [S6]. This is particularly important when the decade box is part of a circuit containing devices that must not have current drawn from them [S6].
Manual Decimal and Sequential Decades Serve Different Needs
Manual decimal boxes are intended for precise resistance substitution, with each dial corresponding to a decimal position and the selected resistor sections combined to produce the total resistance [S2]. A four-dial RZ-4500 therefore follows the units, tens, hundreds, and thousands sequence in its example settings [S4].
Sequential resistance boxes use different step sequences, including 1-2-4 and 1-2-5 arrangements [S2]. A 1-2-5 box can represent 3.6 kΩ with 3,000 Ω and 600 Ω sections, whereas a decimal box represents the same value as 3,600 Ω, 0 Ω, 0 Ω, and 0 Ω across four decades [S2].
Overlap Does Not Raise Accuracy or Power Ratings

The 11th position changes switching mechanics, not the accuracy class or current capacity of a decade. A manufacturer can list an 11-position switch while specifying 0.01% accuracy and separate maximum current and voltage limits for each resistance section [S3][S5].
The IET 1433 specification defines overall accuracy as the difference between resistance at any setting and at the zero setting, equal to the indicated value plus or minus 0.01% and 2 mΩ [S5]. Its published per-step power and voltage limits vary by decade: for example, the 1 Ω step is listed at 1.6 A and 0.16 V, while the 1 kΩ step is listed at 80 mA and 8 V [S5].
These values also show why overlap must not be treated as permission to ignore circuit limits. Sudden, large resistance changes from a decade box may cause excessive and damaging surge currents in the device or system being calibrated and should be avoided [S4].
Selection Depends on the Boundary Transition
Choose an overlap decade when resistance must be changed across a decimal boundary without resetting all dials, and choose a stopped design when accidental direct movement from position 10 to 0 would be hazardous [S6].
Manual decimal boxes fit calibration setups involving instruments such as dial indicators, while resistance substitution also supports sensor and control-loop checks described in Resistance Standards and Decade Steps for Calibration Labs. The panel marking must state each dial’s resistance per step and applicable current rating so the overlap position can be used without ambiguity [S5][S6].
The trackable technical check is the boundary itself: confirm that position 10 overlaps the adjacent decade, identify whether rotation stops there, and verify the stated per-step current and voltage limits before applying the box to a live circuit.