An infrared thermometer needs an emissivity setting because its detector measures infrared radiance, not temperature directly, and the instrument must compensate for how strongly the target surface emits that radiation [S1][S2].
The correction matters most on reflective materials: polished aluminium can have emissivity around 0.02–0.10, while matte black paint is approximately 0.90–0.98, so using the same setting for both can produce materially wrong temperatures [S2].
Emissivity is the conversion factor in an IR measurement
Emissivity, represented by ε, is the ratio of radiation emitted by a real surface to radiation emitted by a perfect black body at the same temperature. A black body has ε = 1.0, while real surfaces lie between the theoretical limits of 0 and 1 [S1][S2].
The measurement chain is straightforward: the surface emits infrared energy, the instrument detects that energy, and the emissivity setting tells the signal-processing electronics how to convert it into a temperature. If the configured value differs from the target’s actual value, the returned temperature is wrong [S3].
This correction is independent of instrument calibration. When an infrared thermometer's emissivity setting does not match the measured material, its temperature reading can be wrong by a margin that invalidates quality, safety, or compliance decisions. For users comparing contact and non-contact methods, Why Contact Thermocouples Are Confused with Infrared Thermometers explains why the two measurement principles respond to different surface and installation effects.
Wrong emissivity produces a predictable error direction
If an instrument is set to 0.95 while the actual surface emissivity is 0.30, the target emits only 30% of the assumed black-body signal. The instrument interprets that weaker radiation as a lower temperature, so the displayed value is too low [S3].
Setting emissivity too low creates the opposite error on a high-emissivity surface, causing overestimation. This behaviour makes a fixed default suitable for many matte, non-reflective materials but unreliable for bare metal, polished surfaces, painted equipment, and targets affected by nearby hot or cold reflections [S2][S6].
Matte black paint has an emissivity close to 0.95, while shiny metals such as aluminum or chrome can have emissivity as low as 0.05.
Surface finish and surroundings can outweigh material labels

Emissivity is not a fixed tag attached to a chemical formula alone. Material, surface finish, oxidation, roughness, viewing angle, contamination, and temperature all affect radiant emission, which is why an emissivity table is a starting estimate rather than a universal property [S1][S5].
A surface's rate of emitting thermal energy depends on its material, surface properties, and temperature, all of which must be accounted for in an infrared thermometer's reading. Infrared thermometers measure infrared emissions from an object, so their readings depend on the object's material, surface properties, and temperature.
Surface condition therefore changes the problem. A weathered or oxidised metal can differ substantially from a clean polished sample of the same alloy, so a material chart alone cannot establish the value for every production component [S5].
Temperature error grows with operating temperature
The Stefan-Boltzmann relationship links total radiated power to ε and the fourth power of absolute temperature, so fractional emissivity errors can produce larger absolute temperature discrepancies at high temperatures [S3].
An infrared thermometer uses its emissivity setting to convert the detected infrared signal into a temperature, and a mismatch between the setting and the material can produce an incorrect reading. These values illustrate the relationship, but exact error depends on instrument band, optics, target temperature, and the direction of the correction.
Typical industrial non-contact thermometers integrate an 8–14 µm waveband, while high-temperature instruments may use a 3–5 µm band [S3]. Wavelength selection can change how strongly a particular material, coating, or viewing condition affects the result, so the emissivity value should be valid for the instrument’s operating band.
Setting the emissivity value in practice

Begin by identifying the actual target surface, including its finish and coating. Select a reference value appropriate to that surface and the instrument’s wavelength band, then enter it using the device controls or configure it through parameters such as USB or RS485 where supported [S5].
A contact comparison provides a stronger field method. Heat the target to a steady temperature, measure the same point with a trusted contact thermocouple, and adjust the infrared instrument until the readings agree. Calex advises against this comparison at room temperature because the adjustment effect is small; it also recommends keeping nearby heaters, furnaces, lamps, and other hot objects out of the reflected field of view [S5] (2020-03).
For bare polished metal, apply matte high-emissivity tape or paint to a small measurement patch, allow it to reach the target temperature, and measure the patch. The emissivity of the tape or coating must be known; painting a surface without accounting for its emissivity only replaces one uncertainty with another [S2].
Compare the main measurement options
Adjustable non-contact instruments offer the widest target coverage because the operator can match emissivity to the material, but they require reliable surface data and control of reflected radiation [S5]. Fixed-emissivity sensors are simpler and work well on non-reflective non-metals or coated metal, yet they are unsuitable for low-emissivity bare metal unless a known high-emissivity patch is added [S5].
Contact probes avoid the emissivity problem because they measure the target through thermal contact, although they can suffer from poor contact, stem conduction, and a disturbance of the measured surface. Infrared devices avoid contact and can inspect large or moving surfaces rapidly, but they depend on a valid emissivity value, a clear line of sight, and control of reflections [S4][S6].
For temperature monitoring and logged process records, a contact RTD PT100 may be preferable where probe placement and thermal contact can be maintained. Where non-contact measurement is required, an infrared thermometer should have adjustable emissivity, and the operating procedure should record the selected value alongside each measurement.
Measurement uncertainty still requires judgment

Correct emissivity does not remove every error source. A dirty lens, partial obstruction of the field of view, reflected radiation, spot-size mismatch, unstable surface temperature, and poor sensor alignment can all degrade the result [S1][S5].
Critical applications require a defensible method rather than a visually estimated coefficient. Record the target material, surface finish, selected emissivity, instrument waveband, ambient temperature, and reflected surroundings; for disputed results, compare against a trusted contact measurement or a suitable reference target [S2][S5].
The practical acceptance check is straightforward: repeat the measurement under stable conditions, compare it with an independent method, and investigate any disagreement before using the value for quality control, maintenance, or safety decisions. A traceable emissivity setting makes the measurement repeatable, but only a verified surface condition makes it defensible.