Class 1 sound level meters require tighter accuracy tolerances and wider frequency coverage than Class 2 meters, making them the stronger choice for laboratory measurement, legal evidence, and defensible environmental surveys [S1][S3].
Type 1 and Type 2 are the older names for Class 1 and Class 2 under IEC 60651, while IEC 61672 uses the term Class and imposes more demanding accuracy, performance, and calibration requirements [S2][S5].
Accuracy class and measurement chain
The accuracy class describes the permitted performance of the complete sound level meter system, not merely the microphone capsule. IEC 61672 tolerance requirements extend across the acquisition chain, and compliance also considers the user interface and the format and detail of recorded results [S2][S4].
One reference-condition tolerance table gives ±0.7 dB for Type 1 and ±1.0 dB for Type 2 under IEC 60651 section 9.1, but these single-level, single-frequency figures do not represent every operating condition [S2].
For procurement, specify the complete standard, class, frequency range, acoustic configuration, microphone, windscreen, calibrator, and required measurements rather than treating one headline tolerance as the complete accuracy claim [S1][S4].
Frequency response and documented bandwidth
Both classes can use A, C, and Z frequency weightings, so weighting selection does not determine class by itself [S1]. The choice depends on required tolerance, usable bandwidth, documentation detail, and whether the result will be challenged [S1][S4].
Selection criteria for each class

Class 1 is for precision work: laboratory and calibration measurements, professional acoustic analysis, environmental assessments requiring defensible data, legal or regulatory reporting, and low-level environmental surveys [S1][S2].
Class 2 is for general field measurements, spot checks, basic surveys, occupational-noise checks, and cost-sensitive monitoring where tighter precision does not affect the decision [S1][S2]. Workplace noise measurement under OSHA or NIOSH guidance can use Class 2 instruments, while some municipalities or governing agencies specifically require Class 1 compliance for event measurements [S4].
A Class 1 meter is not automatically required for every compliance task. Check the applicable regulation, jurisdiction, and contract before ordering.
Class 1 and Class 2 compared
Class 1 systems are required to provide more detailed documentation than Class 2 systems [S4].
Class 2 wins on acquisition cost and suitability for routine surveys, since it meets looser tolerances across a narrower useful bandwidth [S1][S3]. It is the more practical option for general industrial, commercial, educational, recreational, and residential sound-level checks where the result does not require laboratory-grade defensibility [S4].
Dynamic range must be checked separately from class. Class alone does not guarantee adequate range, low self-noise, microphone sensitivity, or the required frequency weighting.
Calibration, features, and failure risks

Do not infer total measurement uncertainty from the class label alone. IEC 60651 reference tolerances cover only one level and one frequency, while the full measurement chain includes microphone response, windscreen correction, interface tolerances, displayed results, and calibration status [S2][S4].
Common failure modes include using a Class 2 meter where the governing rule requires Class 1, recording a measurement outside the instrument’s upper level limit, losing valid data because the selected range is too narrow, and treating a single-frequency tolerance as proof of accuracy across the full measurement chain [S2][S4].
Calibrator compatibility, periodic calibration, windscreen use, microphone orientation, environmental limits, and the required statistic such as Leq should be confirmed before deployment. The A-, C-, and Z-weighting options and optional FFT or octave-band functions do not upgrade a meter from Class 2 to Class 1 [S1].
Purchase and compliance checklist
Specify IEC 61672 class and standard edition in the purchase order, then require evidence that the supplied configuration meets the class rather than relying on a distributor statement. Older IEC 60651 and IEC 60804 instruments may remain usable in many applications, but new procurement should use IEC 61672 because it makes more demanding accuracy, performance, and calibration provisions [S2].
Record the intended application, jurisdiction, required frequency range, expected sound pressure range, frequency weighting, statistical outputs, documentation format, microphone and windscreen configuration, and calibrator model [S1][S4]. For legal evidence, environmental audits, laboratory work, or disputed results, select Class 1. For routine workplace and general field surveys, Class 2 is usually the cost-effective choice when the governing requirement permits it [S2][S4].
The next trackable check is the governing document: confirm whether it names a class, standard edition, tolerance, measurement configuration, and calibration requirement. That comparison should be completed before comparing instrument price, because a lower-cost meter that fails the specified class is not a valid option.