Radius charts are especially well suited for inspecting small-radius form tools, gages, and templates, while 180° protractor charts provide rapid and accurate angle inspection at any desired magnification setting.
An optical comparator can measure a part's shadow at magnifications of up to 100x.
Magnification and projected radius
The working rule is direct: projected radius in millimetres equals the true radius multiplied by magnification. For a 0.05 mm radius, 10x produces 0.50 mm, 20x produces 1.00 mm, 50x produces 2.50 mm, and 100x produces 5.00 mm [S3][S6].
The diameter is twice the radius, so the projected diameter at 20x is 2.00 mm, not 1.00 mm. This distinction matters when aligning a radius template or reading a chart whose dimensions refer to diameter. The projected profile can be compared with screen reference lines, chart overlays, or stage measurements after the image has been focused and oriented [S2].
Magnification does not establish measurement accuracy by itself. A 100x image can be difficult to interpret if the feature is poorly aligned, the illumination is uneven, or the lens and screen produce distortion. The optical comparator must first present a sharp, measurable silhouette.
Selecting between 10x, 20x, and 50x
The explicit comparison is therefore: 10x maximizes field of view, 20x provides a practical 1.00 mm radius image, 50x improves small-radius visibility at the cost of a smaller field, and 100x is available on some comparators for highly detailed work [S2][S5][S6].
Matching the radius chart to the feature

Radius and angle charts are designed to inspect radii with arcs up to 180°, and 180° protractor charts are suitable for any desired magnification setting.
A chart designed for 20x would be the direct match for the 1.00 mm projected radius, while a 50x chart would place the feature at 2.50 mm. The chart must match the comparator setting rather than the drawing radius, otherwise the operator will apply the wrong scale and accept or reject the part incorrectly [S3][S5].
For a small form tool, gage, or template, radius charts are specifically suited to checking small arcs, and 180° protractor charts can support accurate angular inspection without requiring an angle-measuring facility on the comparator [S4].
Tolerance determines whether 20x is enough
Magnification should be tied to the dimensional tolerance, not simply to the nominal 0.05 mm radius.
For example, if the radius tolerance is tight, 50x may be justified because it expands the tolerance band by 2.5 times relative to 20x. If the tolerance is generous and the purpose is a quick production check, 20x may provide enough projected separation without sacrificing the working field of view. The drawing tolerance, surface finish, edge condition, and repeatability requirement determine the final choice [S2][S5].
A nominal radius alone does not support a universal magnification rule. The practical decision is to use the lowest magnification that clearly separates the acceptable and unacceptable profiles while leaving enough field for orientation and stage travel.
Limitations of visual projection

Optical comparators inspect visible 2D geometry, including edges, angles, radii, widths, lengths, and profiles, but they do not recover hidden surfaces or full 3D form [S2]. If the radius is affected by surface texture, a broken or partially obscured edge, uneven illumination, or a nonuniform boundary, a larger image may still produce an unreliable result.
Illumination is central to silhouette work because the comparator projects a shadow or image through a lens onto a screen. Fine burrs, surface waviness, or a dark surface can change the apparent boundary and therefore the apparent radius. Repeatable acceptance requires controlled illumination, a stable focus, correct orientation, and a chart matched to the magnification [S2][S3].
When the geometry cannot be represented cleanly as a silhouette, a coordinate measuring machine or tactile measurement method is the better route. The cited magnification-specimen study discusses measurement methods, comparison of results, and conclusions.
Setting and validating the comparator
A practical inspection starts with a suitable radius chart, clean stage, controlled illumination, and a calibrated magnification setting. Align one reference edge, zero the stage, select the chart scale matching the objective lens, and verify the known radius before checking production parts [S2][S3].
Calibration should establish more than image size. Although that arrangement is not an optical comparator, it illustrates the metrology principle: the measurement scale and its traceability must be verified rather than assumed from the lens marking.
For repeat inspection, use a dimensionally stable chart with sharp, uniform lines, document the objective magnification, and establish an operator acceptance rule. Chart design information identifies the part drawing, available magnifications, illumination, fixturing, inspection rate, and comparator model as inputs for defining the gaging method [S3].
Decision rule for a 0.05 mm radius

Use 20x as the default inspection magnification for a 0.05 mm radius: the radius appears as 1.00 mm and the diameter as 2.00 mm. Move to 50x when the tolerance band, edge condition, or visual resolution requires a 2.50 mm radius image [S3][S5].
Before accepting the setup, measure the projected radius against the selected chart and confirm that the known feature resolves clearly. Track the chosen objective magnification, chart part number, projected radius, and acceptance tolerance for the part. If 20x does not make the tolerance distinction visually unambiguous, move to 50x rather than interpreting an indistinct edge more precisely than the measurement system can support.