A 400 mm casting usually lands somewhere around 150 to 300 µm per voxel, unless regions of interest are scanned at higher magnification.
The attainable voxel size depends on detector pixel pitch, geometric magnification, and field of view, with system geometry, detector characteristics, and part size interacting together rather than part size acting alone.
How Part Size Determines the CT Voxel Floor
For a part rotated through 360 degrees, the entire part must remain inside the detector field of view. The maximum geometric magnification is therefore approximately Mmax = D/d, where D is detector width and d is the widest part dimension. Since voxel size V equals detector pixel pitch P divided by magnification M, the best full-part voxel size approaches Vmin = P × d/D [S3].
A practical sizing rule is V ≈ largest part dimension divided by detector width in pixels. This relationship gives a useful planning estimate, but the final value also depends on focal spot size, reconstruction settings and the selected region of interest [S2][S3][S5].
The industrial CT voxel is a volume element containing reconstructed density information. It is not identical to true spatial resolution, because focal-spot blur, detector response, reconstruction, noise and artifacts can reduce the smallest feature the system can distinguish [S1][S2].
Expected Voxel Sizes Across Part Scales
These figures describe representative operating scales, not guaranteed specifications. A 20 mm component is not automatically a few-micrometre scan, and a 400 mm part cannot inherit the voxel size of a micro-CT configuration designed for a small specimen [S2][S5].
Magnification, Detector Pitch and Focal Spot Limits

Detector pixel pitch P establishes the physical sampling limit, while geometric magnification M converts that detector pitch into the object-space voxel size. Magnification is M = FDD/FOD, using focus-to-detector distance and focus-to-object distance, giving V = P/M [S3].
Moving the part closer to the source increases magnification and reduces voxel size, but excessive magnification makes the source focal spot the limiting blur source. Microfocus X-ray sources address that constraint by minimizing focal-spot size, allowing fine voxels without geometric magnification becoming unrealistically large [S2].
Detector choice is therefore a trade-off among field of view, throughput and spatial sampling, not simply a ranking by pixel count [S2].
Small Parts Versus Large Castings
Small parts support higher magnification because a narrow object occupies less of the detector width. The full-part field-of-view constraint can then be expressed directly through Vmin = P × d/D, linking voxel size to the ratio of detector pixel pitch, part width and detector width [S3].
Large castings reverse that advantage. A 400 mm casting at 150–300 µm voxels provides broad internal coverage, while a higher-resolution scan of a local region may require multiple scans, overlap or a separate workpiece setup [S5].
For dense metal qualification, industrial radiography can remain faster and less costly than full-volume CT when every point in a large casting does not require a reconstructed voxel. CT gains value where volumetric defect maps, wall-thickness measurements and CAD comparison are required [S5].
Defect Size and Measurement Reliability

A feature must generally span several voxels before it can be identified and sized reliably. A pore spanning only 2–3 voxels is a common practical detection threshold, and an industrial industry expectation places surface location within about one-third of a voxel [S5][S6].
This creates a direct acceptance rule: if the smallest defect that must be reliably found is 30 µm, a 10 µm voxel is more defensible than a 20 µm voxel because the defect occupies 3 voxels rather than 1.5. A closed crack narrower than the voxel size, particularly in a large dense part, may remain unresolved [S2][S5].
Feature orientation also matters. Cracks and delaminations are most visible when open enough and favorably oriented relative to the voxel grid. Beam hardening and metal streaks can obscure flaws next to dense inserts, so voxel size alone does not guarantee detectability [S5].
Which CT Configuration Fits the Inspection?
Micro-CT or nano-CT is the correct class for small specimens, internal microstructure, powder particles and micron-scale features where scan time and specimen size are secondary. Laboratory systems support resolutions from 5–150 µm, with specialized nano-CT configurations reaching approximately 0.5 µm [S1][S4].
Industrial cone-beam CT is better suited to castings, additively manufactured components, electronics and assemblies requiring internal geometry over a larger volume. A 20 mm connector can run at a few micrometres, whereas a 400 mm casting typically falls into the 150–300 µm range unless local magnification is used [S5].
The decision can be reduced to four criteria: smallest required defect, largest part dimension, required volume coverage and permitted scan time.
Inspection Boundaries and Practical Thresholds

CT is weak where tight closed cracks fall below voxel size, dense parts exceed available X-ray penetration, and beam-hardening or streak artifacts obscure internal features. Fine surface finish is better handled by optical or 3D scanning, while CT is stronger for inaccessible internal voids, wall thickness and assembly defects [S5].
An overview scan at 150–300 µm voxels does not prove that a 20 µm flaw is absent anywhere in a 400 mm casting. Reliable acceptance requires either a voxel size compatible with the specified flaw and several-voxel sampling rule, or a documented region-by-region strategy at the required resolution [S2][S5].
For a specific job, calculate Vmin = P × d/D from the proposed detector and part envelope, compare that result with the detector pitch and source focal spot, and verify that the smallest acceptance defect spans at least 2–3 voxels. If it does not, use a higher-resolution local scan or a complementary inspection method rather than accepting an under-sampled reconstruction.