An impact hammer is a measuring instrument built around a hand strike. It is also called a modal hammer, an instrumented hammer or an impulse hammer. A force sensor sits in the head behind an interchangeable tip, so every blow on a structure produces an electrical signal proportional to the force that actually entered it. Accelerometers fixed to the structure record the vibration that follows. The ratio of the response to the force, calculated frequency by frequency, is the frequency response function, and from a set of those functions an analyst reads the natural frequencies, the damping and the mode shapes of the structure. Nothing is bolted to the test object and no shaker is needed, which is why the method reaches parts that no other exciter could be attached to.
The method is known as impact testing or impact excitation, and it is one of the two standard ways to excite a structure for experimental modal analysis, the other being an attached electrodynamic shaker. It needs little hardware: the hammer, one or more accelerometers, a signal conditioner or an analyzer with constant-current sensor supplies, and software that computes the frequency response. In return it demands operator skill, because the strength, direction and cleanness of each strike vary from blow to blow. A result is only as good as the input that produced it, and the input here is a hammer blow. The hammer is therefore chosen, set up and checked with the same care as the accelerometers and the analyzer around it.
Photo: NASA Armstrong Flight Research Center / NASA/Steve Freeman, public domain, via Wikimedia Commons
The buying decision concentrates on three matched numbers: the force range, the sensitivity in mV/N and the head mass, because together they decide which structures the hammer can excite without overloading or double striking. The tip set and an extender mass set the frequency band. After that the purchase is judged on its calibration certificate, the durability of its cable and connector, and how well it suits the analyzer that will supply and read it. A hammer bought for one structure is often the wrong hammer for the next one.
Chapter 1 / 06
What an Impact Hammer Is and How It Works
Impulse excitation means that a structure is set in motion by a short force pulse instead of a continuous drive. The hammer delivers that pulse and measures it at the same time. The signal from the head is the input, the signal from each accelerometer is an output, and the analyzer divides the Fourier transform of every output by the Fourier transform of the input at each frequency line. With acceleration as the response the result is accelerance; integrated to velocity it is mobility, and integrated again it is dynamic compliance. ISO 7626-5, the international procedure for these measurements, names all three and covers both driving point measurements, where force and response share a point and a direction, and transfer measurements between different points.
A short pulse is a wide band in the frequency domain, and that is the whole reason the instrument works. The hammer impulse is close to a constant force over a broad frequency range, so one strike excites every mode inside that range at once and there is no need to sweep a frequency as a shaker test must. The spectrum follows from the width of the pulse: the shorter the contact, the higher the frequency at which the energy starts to fall. Striking harder is not the way to widen the band. A harder swing mainly raises the force level, and it wears the tip faster. The frequency at which the content dies away is fixed by how long the tip stays on the surface.
Tip stiffness sets the bandwidth. The frequency content of the blow depends on the stiffness of the two surfaces in contact, the tip and the test object, and to a lesser degree on the mass of the hammer. The object is given, so the test engineer works on the tip. The harder the tip, the shorter the pulse and the higher its frequency content; a softer tip spreads the same momentum over a longer contact and moves the energy toward the lower frequencies. For that reason a hard tip such as steel serves small stiff parts, while a soft plastic or rubber tip serves large compliant ones. As a rule of thumb the tip is chosen so that the force spectrum is no more than 10 to 20 dB down at the highest frequency of interest.
Inside, the head is a stack. The tip screws onto the front of the force sensor. The sensor is a piezoelectric element, commonly quartz, that generates a charge proportional to the force passing through it. Behind the sensor sits the head mass, the inertial block that gives the blow its momentum, and on most hammers an extender mass can be threaded onto the back of that block. The assembly is bonded and locked at the factory and is not meant to be opened. Because the tip lies in front of the sensing element, the sensitivity belongs to one hammer configuration, and a different tip or an added extender mass is a reason to recheck it.
Two electrical designs are common. A charge-mode hammer sends the raw charge of the crystal down a low-noise cable, a shielded cable with a treated dielectric, to a charge amplifier that turns it into a voltage; it is the form used where the temperature exceeds what built-in electronics accept. An IEPE hammer, one with integrated electronics, has a unity-gain amplifier built into the sensor that outputs a low-impedance voltage, and its built-in amplifier standardizes the sensitivity within a few percent of the nominal value, which is adequate for most dynamic applications. It needs a constant-current supply, 2 to 20 mA from a 20 to 30 VDC source, which a signal conditioner or many analyzers provide, and it runs over an ordinary coaxial cable. A data logger for slow process channels cannot capture the pulse.
What physically happens is an exchange of momentum. The head arrives with mass and velocity, the tip compresses against the surface, and a force pulse rises and falls within the contact time. By the law of linear momentum the impulse equals the change of momentum of the head, so a heavier head or a faster swing gives a larger force. The velocity of a hand swing is hard to control, so in practice the force level is set by choosing the head mass and by adding or removing an extender mass. The structure answers with a decaying vibration that is the sum of its modes, and the analyzer captures both the pulse and that decay in one timed record.
Chapter 2 / 06
Impact Hammer Types and Structural Variants
Most impact hammers are hand-held manual hammers in the shape of an ordinary hammer: a cylindrical head of 6.3 mm (0.25 in) to 76 mm (3.0 in) in diameter on a handle, with the cable leaving the head or the end of the handle. At the small end a head of 4.8 g (0.17 oz) with a sensitivity of 22.5 mV/N is made for circuit boards, processors and memory modules. Mid sizes around 0.16 kg cover panels, frames, small electric motors and engines. The largest hand-held forms are sledgehammers of about 5.5 kg (12.1 lb) for large weldments, propellers and building foundations. The product class runs from 4.8 g to 5.5 kg in head mass and from 222 N (50 lbf) to 22,240 N (5,000 lbf) in peak force range.
Heavier forms leave the one-hand format. For large, heavy and compliant structures such as locomotive frames, tanks and bridges, the instrumented sledgehammer is used, and some very large structures need a massive mechanical ram fitted with a force-sensing impact head instead. The head of such a unit is large, so that its contact area and force range match the structure. Its resonant frequency is lower, at least 5 kHz at the top of the size range against at least 100 kHz for the smallest head, and its flat band therefore ends earlier. Its sensor is hermetically sealed where smaller heads are epoxy sealed. Sensitivity falls as the force range rises: the largest heads give about 0.23 mV/N (1 mV/lbf), the smallest about 22.5 mV/N (100 mV/lbf).
Variants also differ in where the conditioning sits. A charge-mode hammer yields a charge in pC/N and depends on an external charge amplifier, one more box to carry, power and calibrate. An IEPE hammer carries the impedance conversion in its head and plugs straight into an analyzer input that supplies constant current, with no extra signal conditioning. A further step is the hammer with a built-in transducer electronic data sheet, known as TEDS: the head stores its sensitivity and identification, and the analyzer reads them on connection, which removes typing errors when many channels or many hammers are in use. The feature is typically chosen for tests with a high channel count and for roving hammer surveys.
The construction that separates a modal hammer from a force sensor fixed to a handle is modal tuning. A hammer is itself a structure with its own resonances, and a head that rings after the strike would superimpose its own motion on the force signal. A modally tuned head is built so that the hammer's structure does not influence the measurement, and the rigid body calibration described later shows at once whether a head is tuned, because a hammer that is not tuned distorts that result. Tuning belongs to the bonded assembly of sensor, head mass and handle: the sensor element is laser welded, the assembly is locked together with a structural adhesive at the factory, and it is not taken apart except at the factory. The parts the user changes are the tip and the extender mass. The tip transfers the force of impact to the sensor and protects the sensor face, and tips of different stiffness vary the pulse width and the frequency content of the force.
The word hammer causes most of the confusion. A demolition hammer and a rotary hammer are motor-driven tools that break or drill material, and a hammer mill is a machine that crushes by impact; none of them measures anything. An impact hammer is a force transducer with a handle. It sits in the same family as the load cell, a force sensor used for sustained and static loads, and the force gauge, a hand-held indicator for steady pull and push. Inside such a load cell, a strain gauge measures force through the bending of a spring element, while the piezoelectric element, the type most widely used for modal testing, generates a charge directly and covers wide frequency and dynamic ranges with good linearity. The instrument is not complete on its own: a vibration analyzer supplies, captures and processes the signals, and the response sensor is its other half.
Chapter 3 / 06
Impact Hammer Key Specification Parameters and Standards
Sensitivity, force range and head mass are the headline figures, and they trade against each other. Sensitivity is the output voltage per unit of force, quoted in mV/N or mV/lbf with a tolerance of plus or minus 15 percent. It is chosen so that the force the structure will receive uses most of the analyzer input range but not all of it; a hammer worked in the upper half of its range leaves no room for the variation between strikes. The force range is the peak force the sensor measures, from about 222 N (50 lbf) in the smallest heads to 22,240 N (5,000 lbf) in the largest. Resonant frequency is the natural frequency of the sensor assembly, at least 5 kHz to at least 100 kHz, and the test band must lie below it. Non-linearity is held to 1 percent or less.
The usable frequency range is not a fixed number on a datasheet, because the tip sets it. The same head covers a far higher band with a steel tip than with a rubber tip, and the extender mass, threaded onto the back of the head, lowers the band while raising the force delivered. Suppliers publish a force spectrum for each tip, and the purchaser reads the highest frequency at which the curve is no more than 10 dB to 20 dB down. The electrical figures belong to the built-in electronics of an IEPE head; for one hammer line they are a constant-current excitation of 2 to 20 mA, a supply of 20 to 30 VDC, an output bias of 8 to 14 VDC and a discharge time constant of 100 s or more, which sets the lowest frequency the sensor can measure down to.
Three standards matter, and each governs a different part of the purchase. ISO 7626-5 is the international standard for measuring mechanical mobility, accelerance and dynamic compliance with an exciter that is not attached to the structure, which is the procedure the hammer exists to serve; it is written as a measurement procedure and not as a product specification for the hammer, and it is limited to accurate mobility measurements, not to quick uncalibrated surveys. ISO 16063-21 governs the calibration of the accelerometers by comparison with a reference transducer over 0.4 Hz to 10 kHz, in the laboratory or in the field; it does not cover the force sensor. ISO/IEC 17025 sets the competence required of the calibration laboratory that issues the certificates.
Parameter
What it states
Typical value or limit
Sensitivity
Output per unit force
0.23 to 22.5 mV/N (1 to 100 mV/lbf), plus or minus 15 percent
Force range
Peak force measured
222 N (50 lbf) to 22,240 N (5,000 lbf)
Overload margin
Force never to exceed
No more than 5 times the rated range
Head mass
Mass of the head
4.8 g (0.17 oz) to 5.5 kg (12.1 lb)
Head diameter
Size of striking end
6.3 mm (0.25 in) to 76 mm (3.0 in)
Resonant frequency
Upper bound of flat range
At least 5 kHz to at least 100 kHz
Non-linearity
Deviation from linear output
1 percent or less
Sensing element
Material of force sensor
Quartz, piezoelectric
Tips
Interchangeable contact faces
Steel, plastic or rubber, very soft to hard
Extender mass
Mass added behind head
1.25 g to 200 g
Excitation (IEPE)
Supply from conditioner
2 to 20 mA, 20 to 30 VDC
Output bias (IEPE)
DC level at rest
8 to 14 VDC
Discharge time constant
Low frequency limit
100 s or more
Temperature limit
Built-in electronics
121 degrees C (250 degrees F)
Chapter 4 / 06
Impact Hammer Selection and Duty Calculations
Choosing an impact hammer is a matching exercise. The structure brings its mass, its stiffness, its size and the frequency band that matters; the hammer brings head mass, force range, sensitivity and a set of tips. The test engineer matches the two with a few rules and one trial measurement before any data is kept.
The first choice follows from the structure. A part of a few grams, such as a circuit board, a processor module or a memory module, takes the smallest head and a hard tip, because a heavy head bounces off it and double strikes. A car frame, an engine or a small electric motor takes a medium head. Pumps, compressors, weldments and impellers are served by heads of 0.16 kg (0.34 lb) to 1.1 kg (2.4 lb), large weldments and propellers by heads of 1.1 kg (2.4 lb) to 5.5 kg (12.1 lb), and building foundations by the heaviest sledge. Within that size the tip follows from the band of interest, and a softer tip with the extender mass fitted is the answer when that band lies low.
The first calculation is the output that the expected force will produce. Multiplying the sensitivity by the peak force gives the voltage: a head with a sensitivity of 2.25 mV/N struck to 2,224 N produces about 5 V, which is why the rated range of such heads corresponds to roughly 5 V at full scale. The analyzer input range is then set at no more than twice the maximum signal, a practice called half ranging, so that the analog to digital converter is used fully without clipping. The same margin is kept on the hammer: working in the upper half of the force range leaves no allowance for one strike that lands harder than the rest, and flat topped peaks are the sign that the chain has saturated.
The second calculation sets the time record. A wide analysis band gives a short time record, and on a lightly damped structure the response is still ringing when the record ends. The result is leakage, an error that lowers the measured amplitude and raises the measured damping. Two windows treat it. A force window removes the noise after the pulse without cutting the pulse itself, and an exponential window forces the response to decay inside the record, with a time constant of about one fourth of the record length. That second window adds artificial damping, which the software removes again, so it runs into numerical trouble where the added damping far exceeds the true damping. For those structures a zoomed band with a longer time record is the better choice.
The third calculation is the number of measurements. A row or a column of the frequency response matrix is enough to identify all modes, so the two standard arrangements are equivalent in principle for a linear structure. With a fixed accelerometer and a roving hammer the hammer is walked over the grid of points and the structure is struck at each in turn; with a fixed excitation point and roving accelerometers the sensors are moved instead. Either way one measurement must be a driving point, with force and response at the same point and in the same direction, or the mode shapes cannot be scaled. Each point is averaged over at least 4 impacts, the capture is triggered on the force channel with a pre trigger so that the whole pulse is inside the block, and every strike is inspected before it is accepted.
Quality is judged from three displays. The coherence function shows, frequency by frequency, how much of the response is linearly caused by the measured force; it collapses where the response is too weak, where an input has clipped, or where a second strike has corrupted the record. The force trace shows a double hit as a second pulse, and such a strike is rejected rather than averaged. In the imaginary part of a driving point measurement every resonance peak must lie in the same direction; one that points the other way means a reversed sensor direction or two different points. Mass loading is checked by mounting a second accelerometer of equal mass at the same point and repeating the measurement: if the peaks move, the response should be taken with a lighter accelerometer or without contact, by a laser displacement sensor or a proximity probe.
Impact Hammer Installation, Maintenance, and Typical Failure Modes
An impact hammer has no installation in the sense of a bolted instrument. Its setup is the tip and extender mass that are fitted, the cable that is run, the supply it is connected to and the checks made before the first strike. The steps are short, and skipping them is the usual beginning of a bad data set.
Fitting is by hand. The tip and the extender mass thread onto studs at opposite ends of the head, and both are tightened firmly; repeated impacts loosen them and produce erratic, noisy signals. The cable runs from the end of the handle to the conditioner or analyzer, and its connectors are tightened by hand. Enough slack is left for the swing, because the cable adds effective mass to the handle, most noticeably on the smallest heads, and on large hammers the cable is made the weak link so that a missed hit breaks the cable rather than the sensor. The analyzer is grounded to keep interference from fluorescent lighting and other sources out of the signal.
After the supply is switched on, 1 to 2 minutes pass before the amplifier has settled and the coupling capacitor has charged. A digital multimeter or an oscilloscope across the sensor output then reads 8 to 14 VDC, the bias of an IEPE hammer at rest, which confirms that the electronics are alive and the cable continuous; a value near zero or near the supply points to a short or an open circuit. Each accelerometer is mounted so that the highest test frequency lies below one tenth of its mounted natural frequency. The structure is supported but not constrained, because a support adds damping, and where a free test is wanted the support is soft enough that the highest rigid body mode lies below one tenth of the first flexible mode. Practice swings follow, aimed at the most direct impact with the least chance of a second contact.
Calibration of the hammer is a dynamic check that needs no dead-weight force standard. The hammer is struck against a mass that is suspended like a pendulum or resting on soft foam, so that it moves as a rigid body, and a reference accelerometer on that mass measures its acceleration. By Newton's second law the force equals the mass times the acceleration, so the peak voltage of the hammer divided by the product of mass and peak acceleration gives the sensitivity in mV/N. The reference accelerometer forms part of the traceability chain: it is calibrated by comparison under ISO 16063-21, and a laboratory accredited to ISO/IEC 17025 issues the certificates. Recalibration is recommended every year, after any extreme environmental exposure and before a critical test.
Handling rules protect the sensor face and the built-in electronics. A tip is always fitted, because the lapped face of the sensor is damaged by a bare strike and its behavior changes afterwards. The hammer is never driven beyond 5 times its rated force range, since a larger force destroys the miniature electronics, and no voltage is applied without constant current protection, with 20 mA and 30 V as the limits. The temperature stays below 121 degrees C (250 degrees F). A hammer that has been dropped is checked for a cracked tip, crossed threads and a bent connector before it returns to service. Cables are replaced at the first sign of intermittent output, since they are often the source of trouble. Connectors are wiped with a damp cloth, and a unit that is not hermetically sealed is never soaked or submerged.
Symptom
Likely cause
Action
No output, bias near 0 V
Shorted cable or connection
Replace the cable
No output, bias near the supply
Open cable or connection
Reseat, then replace the cable
Flat topped force peak
Sensor or input saturated
Strike more lightly or use a larger range head
Second pulse on force trace
Double hit, head too heavy
Reject the strike, use a lighter head
Ripple on force spectrum
Bouncing or penetration at contact
Reject the data, change tip or head
Force spectrum rolls off early
Tip too soft
Fit a stiffer tip, remove the extender mass
Little low frequency response
Tip too hard or head too light
Softer tip, extender mass, heavier head
Erratic, noisy signal
Loosened tip, extender or cable
Check and tighten every connection
Slow negative dip after pulse
AC coupled input
Set the input to DC coupling
Chapter 6 / 06
Impact Hammer Industry Applications and Procurement Checks
Impact testing is used in every industry that owns a resonant structure. Aerospace and vehicle laboratories strike frames, engines, small electric motors and complete assemblies to confirm the modes that their finite element models predict. In process plants the same method finds the casing modes of a centrifugal pump, a centrifugal compressor or a gearbox, where a natural frequency close to a running speed or a gear mesh frequency turns small forces into noise and fatigue. A centrifugal fan is tested at rest so that its blade passing frequency can be kept clear of a natural frequency. On machine tools the hammer is applied at the tool point while an accelerometer sits in the spindle, the measurement that predicts the depth of cut at which chatter begins. Robot builders strike the links of an articulated robot to find the modes that limit accuracy at speed.
In maintenance the hammer is a diagnostic tool. A machine that vibrates at one speed and not at another is measured at rest, and a natural frequency found near the running speed explains the peak in its spectrum; the cure is then a change of speed, a stiffer mount or added mass, and a tachometer confirms the speed once the machine runs again. A dynamic balancing machine has no part in this, because unbalance is only the exciting force and the resonance is what amplifies it. At the other end of the scale, makers of circuit boards and precision components test samples that weigh a few grams, and the panels of a machine noise control housing are struck to find the ring frequencies that let noise through.
The method has limits that a purchaser should understand before relying on it for every structure. The energy of a hand strike is small, so a very large or heavily damped structure can respond too weakly; an attached shaker then takes over, driven by a function generator through a power amplifier. The force spectrum cannot be band limited at the low end, so a zoomed measurement still excites the modes below the band of interest. The result depends on the operator. Where the aim is only to watch a machine in service, a vibration sensor fixed to the housing and read by a condition monitoring system is the right instrument. A vibration meter that shows the overall level of a running machine has no use in a modal test, and the hammer has no use in route-based monitoring.
An order for an impact hammer states the structure first and the instrument second. Give the mass range and the size of the structures to be tested and the frequency band that matters, and let the supplier confirm the head mass, the force range and the sensitivity in mV/N with their tolerances. Specify the tip set, from hard steel to soft rubber, and whether an extender mass is included. State the electrical interface: IEPE output with a coaxial connector for direct connection to the analyzer, or charge output with a charge amplifier. Ask for the force spectrum for each tip and head combination, the overload limit and the temperature limit. Require a calibration certificate with the sensitivity of the delivered configuration, the method used, the measurement uncertainty and the name of a laboratory accredited to ISO/IEC 17025; where accelerometers are bought with the hammer, their certificates should cite ISO 16063-21. If the measurements must stand as mobility data, name ISO 7626-5 in the test procedure. Add spare tips, a spare cable and a reference mass for the field check, and ask for the recalibration interval and the repair path.
FAQ
How does an impact hammer work?
An impact hammer has a force sensor built into its head behind an interchangeable tip. When the tip strikes a structure, the sensor produces a voltage proportional to the force that enters it, while one or more accelerometers on the structure record the vibration. The analyzer divides the response spectrum by the force spectrum at every frequency line and so builds a frequency response function. Because a single short pulse contains a broad band of frequencies, one strike excites every mode in that band at once. The natural frequencies, damping and mode shapes are then identified from a set of such functions.
What types of impact hammer are there?
Impact hammers are typed by size, by electronics and by extras. Sizes run from a head of 4.8 g (0.17 oz) for circuit boards to a 5.5 kg (12.1 lb) sledgehammer for large weldments and foundations, with peak force ranges from 222 N (50 lbf) to 22,240 N (5,000 lbf). By electronics there are charge-output hammers that need an external charge amplifier and IEPE hammers with built-in electronics that plug into a constant current analyzer input. Some heads carry a transducer electronic data sheet so the analyzer reads the sensitivity itself.
Which specifications matter most on an impact hammer?
Sensitivity, force range and head mass lead. Sensitivity, the output per unit of force, typically runs 0.23 to 22.5 mV/N (1 to 100 mV/lbf) with a tolerance of plus or minus 15 percent. The force range is the peak force the sensor measures, and the head mass decides how much momentum a strike delivers and how likely a double hit is. The resonant frequency of the sensor, at least 5 kHz to at least 100 kHz depending on size, bounds the usable band from above. In practice the tip set sets that band, because a harder tip gives a shorter pulse and more high frequency content.
How is the right hammer and tip chosen?
Start from the structure and the frequency band. Light parts such as circuit boards and small machined components take the smallest head with a hard tip; car frames, engines and small electric motors take a medium head; pumps, compressors and weldments take heads of 0.16 kg to 1.1 kg, and large weldments, propellers and foundations take the heaviest heads. Pick a tip whose force spectrum has dropped by no more than 10 to 20 dB at the highest frequency that matters. For a low band, fit a softer tip and the extender mass; for a high band, use a stiffer tip without the extender. Take a trial measurement. A second pulse on the force trace means the hammer is too heavy for the part.
What causes a double hit and how is it avoided?
A double hit is a second contact between the tip and the structure within the same measurement block. It happens when the head is too heavy for the part, when the swing is not straight, or when the structure rebounds into the hammer. The force trace shows a second pulse, and the spectrum of a double strike carries an oscillatory component superimposed on the data, which is why such a record is rejected and not averaged. Such data is rejected and the point is struck again. To avoid it, use a lighter head, shorten the swing, aim straight at the point, and practice the stroke before capturing data.
How is an impact hammer calibrated and how often?
The usual method is a rigid body test. The hammer strikes a mass that hangs freely or rests on soft foam, and a reference accelerometer on the mass measures the acceleration. Since force equals mass times acceleration, dividing the peak voltage of the hammer by mass times acceleration gives its sensitivity, and an FFT analyzer gives the same result at every frequency. Calibration belongs to the configuration: a different tip mass or a fitted extender mass changes the sensitivity slightly, so the certificate should match the parts in use. Recalibration every year, after any extreme environmental exposure and before a critical test, is the common recommendation.
Which standards apply to impact hammer testing?
Three standards apply to three different things. ISO 7626-5 gives the procedure for measuring mechanical mobility, accelerance and dynamic compliance by impact from an exciter that is not attached to the structure; it serves accurate measurements and is not a product specification for the hammer. ISO 16063-21 covers calibration of accelerometers against a reference transducer from 0.4 Hz to 10 kHz, in the laboratory or in the field; the force sensor is outside its scope. ISO/IEC 17025 defines the competence required of the laboratory that issues the calibration certificates for both instruments.